AQUALAB TDL 2 Precise water activity data for even the most volatile samples.

Volatility meets precision. Achieve accurate water activity readings in samples with volatile compounds.

Categories:

Learn more:
Technical Specification
  • Water Activity Measurement:
    • Range: 0.0000 to 1.0000 aw
    • Resolution: 0.0001 aw
    • Accuracy: ±0.0050 aw at 25 °C
    • Repeatability: 0.0010 aw
    • Sensor Type: Tunable diode laser
  • Moisture Content Measurement:
    • Accuracy: 0.1%–0.5% to moisture content reference*
    • Repeatability: 0.02%
  • Temperature Measurement:
    • Range: 15–50 °C
    • Resolution: 0.01 °C
    • Accuracy: ±0.10 °C
    • Adjustment Increment: 1 °C
  • Read Time: ~5 minutes
  • Physical Specifications:
    • Case Dimensions: 26.7 cm (L) × 17.8 cm (W) × 12.7 cm (H)
    • Case Material: POLYLAC PA-765 (ABS) with fire retardant
    • Display: 64 × 128 graphical
    • Power: 85-264 VAC, 47-63 Hz
    • Sample Cup Capacity: 14 mL (0.47 fl oz)
    • Weight: 3.1 kg (6.8 lb)
  • Operating Conditions:
    • Temperature Range: 4.00 °C to 50.00 °C
    • Humidity: 0%–90% RH (non-condensing)
  • Data Communications:
    • USB A and RS-232 serial (9,600–115,000 baud)
  • Compliance:
    • Manufactured under ISO 9001:2015 and ISO/IEC 17050:2010 standards
    • CE Mark certified

Make an Enquiry